詳細(xì)介紹
Monitors 0.05 μm particle contamination in pure water
Integrated unit combines sensor, controller, flow meter, and data storage
Install directly to pure water line and monitor the process
Purge air unit installed as standard
High pressure up to 500 kPa (gauge pressure) applicable
Equipped with D/A converter (4 - 20 mA) and alarm output contacts (standard)
Printer built-in, CompactFlash card available (option)
Specification
Optical system | Light-scattering method |
---|---|
Light source | Laser diode (wavelength 830nm, rated output 200mW ) |
Materials of parts exposed to sample fluid | Synthetic quartz, fluorocarbon rubber, fluoroplastic,PVC, Pyrex glass, SUS304/316, POM |
Measurement size range | 4 channels (Factory default): ≥0.05µm, ≥0.1µm, ≥0.15µm, ≥0.2µm Setting range 0.05µm to 0.2µm |
Sanpling Flow rate | Flow rate 10mL/min and purge flow rate 0.1 to 1L/min combined (Purge flow rate will differ depending on sample fluid pressure) |
Maximum particle number concentration | 200,000 particles/mL (coincidence loss 10%) |
Power requirements | 100V to 240V AC, 50/60Hz, Approx. 80VA |
Dimensions | Approx. 230 (H) × 330 (W) × 569 (D) mm |
Weight | Approx. 19.8kg |